[SHTM 03-01 Part A – PDF page 158]

A spot reading will be taken at the face of the filter to determine the background particle level. A leak is defined as a significant and repeatable rise above the background level.

Terminal (EPA or HEPA) filter media leak test

The test will confirm that the terminal filters have not sustained damage while being installed.

The face of each terminal filter should be scanned using an LSAPC to prove that there are no leaks.

A leak is defined as a significant and repeatable rise above the background level.

Cleanroom particle count

The test will confirm the number and size of particles present and therefore the classification of the cleanroom in terms of ISO 14644 or EUGGMP as specified in the project brief.

The number of test positions is determined by reference to Table A.1 in ISO 14644-1.

The complete test methodology will be as set out in ISO 14644.

Cleanroom biocontamination control

BS EN 17141 gives details on cleanroom biocontamination control.

Radiopharmacy aseptic preparation facilities

Validation will be as for a pharmacy aseptic preparation facility.

Additional radiological tests as specified in the project brief will be required. These will be carried out and/or witnessed by the client’s appointed specialist.

Inspection, assembly and packing (IAP) rooms in central decontamination units

Validation will be as for the standard practice described in paragraphs 12.1–12.33.

The pressure cascade and associated automatic monitoring sensors and alarms should be tested for correct operation in accordance with the design specification.

Note: The detail of the sealing between the instrument washers, transfer hatches and sterilizers that penetrate the walls of the IAP room will be critical in attaining the specified room pressure.

Following the satisfactory validation, the IAP room should be physically cleaned using specialist contractors. Particle counts at locations related to the floor area as set out in table A.1 of ISO 14644 Part 1, along with instructions from ISO 17141 will then be used to establish whether the room achieves a Class 8 cleanroom standard.

[SHTM 03-01 Part A – PDF page 159]

Containment level 3 laboratories

Validation will be as for the standard practice described in paragraphs 12.1–12.33.

The room will be subject to a permeability test as set out in paragraph 12.17.

The pressure cascade and associated automatic monitoring sensors and alarms should be tested for correct operation in accordance with the design specification.

Isolation rooms

Validation will be as for the standard practice described in paragraphs 12.1–12.33.

See Scottish Health Planning Note 4 Supplement 1 for details of the test regime.

Microbiological sampling

It is essential that all parts of the validation test specified above have been successfully completed and the areas thoroughly cleaned prior to any microbiological sampling.

Microbiological sampling will not normally be required for either general or local exhaust ventilation (LEV) systems unless otherwise specified within the contract.

The procedure for carrying out microbiological sampling in cleanrooms is set out in ISO 17141.

Microbiological sampling conventional theatres

The level of airborne bacteria introduced by the supply air can be checked by closing all doors and leaving the operating room empty with the ventilation system running for 15 minutes. An active air sampler set to 1 cubic metre and mounted on the operating table should then be activated remotely. Aerobic cultures on non-selective media should not exceed 10 bacterial and/or fungal colony forming units per cubic metre (CFU/m3).

The results should be examined to establish the broad category of organisms present. A high preponderance of fungal organisms may be an indication of inadequate filtration for the particular installation. Precise guidance is inappropriate and will depend on local circumstances.

It may be appropriate to carry out a check of airborne bacteria during a surgical operation. If required, this should be carried out as soon as possible after handover. Unless there are unusually high numbers of personnel or extensive activity in the room, the number of airborne bacterial and/or fungal CFU averaged over any five- minute period, would be unlikely to exceed 180 per cubic metre.

Information on the microbiological testing of UCV Operating suites is given in clauses 12.124 and 12.125.