[HTM 03-01 A – PDF page 163]

Appendix 2: Summary of design conditions

Appendix 2: Summary of design conditions

Application Ventilation Air-change

Supply filter

Temp

rate (ac/h)

Pressure (Pascal – Pa)

Noise (dB(A))

(°C)

Comments (for further information

grade (BS EN 16798)

see Chapter 8)

S/N – SUP2 18–28

General ward (level 0 and 1 care)

E –ve – –

Communal ward toilet

Single room S/E/N 0 or –ve SUP2 18–28

Single room WC E –ve – –

Clean utility S +ve SUP3 18–22

Dirty utility E –ve – –

S Lobby +10

Room 0

Ward isolation room (PPVL)

SUP2 – See Health Building Note 04-01 (Supplement 1)

E –5 SUP2 – See Table 4

Infectious diseases isolation room

S +10 H12 – See Table 3

Neutropaenic patient ward

Critical care areas (Level 2 and 3 care)

S +10 SUP1 – Isolation room may be –ve pressure or PPVL. See Table 3

Birthing room S & E SUP2 20–25 See Table 5

NICU/SCBU S & E +ve SUP1 20–28 Isolation room may be –ve pressure

For general and UCV operating suites and associated rooms, see specific guidance in Chapter 8 and typical design solutions in Appendix 7

S & E SUP2 18–25 Provide clean airflow path

Operating department recovery room

S & E +ve SUP2 18–22

Catheterisation room

[HTM 03-01 A – PDF page 164]

Application Ventilation Air-change

Supply filter

Temp

rate (ac/h)

Pressure (Pascal – Pa)

Noise (dB(A))

(°C)

Comments (for further information

grade (BS EN 16798)

see Chapter 8)

S & E +ve SUP2 – Stable conditions as specified for the imaging equipment

Interventional or non- interventional Imaging room of any type

S & E S/E SUP2 18–28

Sedation recovery room as in paragraph 8.16

S & E -5 SUP2 20–25 See Table 2

Endoscopic procedure room

E –ve – –

Endoscope reprocessing wash room

S & E Neutral SUP2 20–25 See Table 2

General treatment room

S & E – SUP2 – 18–25 See Table 2

Emergency department waiting area

Containment level 3 laboratory >20 H14* – 18–22 # See ACDP guide; *Filter in extract See Table 4

Post-mortem room

S & E S = 10 E = 12

–ve SUP2 18–22 Provide clean airflow path

Specimen store E – –ve – – – Fan accessible from outside of store

Notes:

Waiting and circulation areas should be directly or indirectly ventilated to provide a comfortable environment and control airborne contamination and odours.

18–22°C indicates the range over which the temperature may float.

18-22°C indicates the range over which the temperature should be capable of being controlled.

S = Suppy

E = Extract

N = Natural ventilation where possible