HTM 03-01 A: _Hlk37000395 – UCV canopy entrainment test
[HTM 03-01 A – PDF page 146]
UCV 1m air velocity test set-up
are recommenced starting at the right of the test row and working from right to left or a quadrant at a time, as above.
UCV canopy high level discharge velocity test
12.82 Measurements of air velocity are to be taken at every test position 2 m above floor level and the results averaged. The average of the total readings taken is to be not less than:
UCV canopy entrainment test
Rationale for the entrainment test
• 0.38 m/s for a canopy with no side walls or side walls that terminate at 2 m above floor level.
• 0.30 m/s for a canopy with side walls that terminate 1 m above floor level.
12.86 The performance of a UCV canopy may be compromised by room air being drawn into the ultra clean airflow, a phenomenon known as entrainment. Significant levels of entrainment could lead to microbial contamination of items left exposed on instrument trolleys laid out beneath the canopy.
12.83 For UCV canopies that are an assembly of two or four units, each fed by a recirculation fan, the average air velocity for each unit should not exceed ±6% of the measured average velocity for the canopy.
UCV 2m air velocity test set-up
12.87 UCV canopies having permanently fitted side walls that terminate 1 m above floor level do not need to be tested, as the walls physically prevent entrainment.
Principle of the test
UCV canopy low level air velocity test
12.84 Measurements of air velocity are to be taken at each of the inner zone test positions 1 m above floor level.
12.85 The measured velocity at every test position in the inner clean zone should be not less than 0.20 m/s.
12.88 A source of particles is produced outside of the UCV canopy footprint and is used to challenge the system. A sample probe and detector are placed within the ultra clean airflow and used to determine the percentage penetration of the test particles at predefined locations under the UCV canopy footprint. The source and sample probe are moved in tandem around the UCV canopy and pairs of readings taken at the detector, from which the percentage penetration at specified locations is calculated. The degree of penetration should be below specified maximum limits if entrainment is to be declared not significant.
[HTM 03-01 A – PDF page 147]
12.0 Acceptance testing: validation
UCV entrainment test setup
Test setup
12.89 The challenge will be provided by using non-EPA-filtered air emerging from the preparation room via the pressure stabiliser or transfer grille and ducted to the specified release position.
12.90 The canopy face diffuser screen should be in place for these tests.
12.91 The test is performed without any theatre equipment in place beneath or closely adjacent to the UCV canopy. All doors in the theatre suite should be closed and remain so for the duration of the test.
12.92 The operating lights and support booms should be moved to a central position beneath the canopy and raised to 2 m above floor level, so as not to interfere with the peripheral airflows (see photograph).
12.93 Spot readings are taken at the centre of the canopy, 1 m from floor level, to establish that the room is within the specified temperature and humidity limits (see paragraph 12.47).
12.94 The test grid is set out as described previously (see paragraph 12.50).
Test equipment
12.96 The detector will be an LSAPC capable of sampling a minimum of 28.3 L of air (1 ft3) per minute and providing readings for particle sizes from to 5 µm to 0.3 µm. The instrument should be compliant with the requirements of BS EN ISO 14644. An alternative instrument may be used providing it is of no lesser specification.
12.97 The sampling head will be an isokinetic fishtail scanning probe mounted horizontally on a test stand 1 m above floor level and connected to the LSAPC by a hose no longer than 2 m.
12.95 The source unit will be a fan/blower or other method that ducts non-EPA- filtered air (see paragraph 12.89) and expels it via a delivery head mounted on a test stand or clamped to the UCV canopy sidewall at the specified release position to provide the particle challenge. The challenge air will be delivered vertically downwards from a position 2 m above floor level alongside the outside edge of the side wall or in line with the downward air curtain if the canopy does not have side walls. The challenge airflow velocity should be the same as the measured average velocity at the 2 m level for the canopy under test.
[HTM 03-01 A – PDF page 148]
no less searching examination of the unit’s ability to control entrainment.
Test positions and orientation of source and detector sampling probe
Test method
12.98 The test positions will be at the centre of each test square, as defined for the velocity test (see paragraph 12.50).
12.103 A measurement of particle penetration through a representative section of the EPA filter media is to be taken. The smallest non-penetrating particle size will be used as the reference background level and set in the detector instrument. The detector instrument should be set to take a reading over a 15-second sample interval and record the number of particles at the non-penetrating particle size determined above.
12.99 For rectangular UCV canopies, measurements of penetration are to be taken at the four corner test squares of the test grid and at intermediate positions along the line of test squares between the corners. The number of intermediate test positions will be as equally spaced as possible around the periphery, with not fewer than three and not more than five complete test squares between test positions.
12.104 An initial sample of air at the source delivery head should be taken to check that there are sufficient particles of the considered size present. The challenge will be considered suitable if:
a. the particles are within the size range
5 to 0.3 µm and thus capable of remaining airborne for a substantial time.
b. the particles should not be able to
12.100 A further series of measurements are to be obtained around the periphery of the inner zone (defined in paragraph 12.54). Measurements of penetration are to be taken at the four corner test squares of the inner zone of the test grid and if necessary at intermediate positions along the line of test squares between the corners as equally spaced as possible, with not fewer than three and not more than five complete test squares between test positions.
penetrate the canopy EPA filters in sufficient numbers to cause a background count that is more than 0.1% of the challenge count.
c. the number of particles present will
enable a minimum of three logarithm (1000-fold) range of counts to be recorded between the source and background readings. A concentration of approximately 105 particles per cubic metre of source air has been shown to be adequate.
12.101 The centre of the challenge particle source delivery head is aligned with the centre of the designated test square, with its longer edge against the outer edge of the side wall or air curtain and delivering the challenge 2 m above floor level. The air containing challenge particles is directed vertically downward. Where there is physical interference due to obstructions such as gas pendants, the source will be moved to the next available non-obstructed test square location nearest to the stipulated test position. The sampling probe will then also be moved to remain opposite the source.
The same equipment should be used to measure both the challenge source and penetration so as not to bias results through particle losses within the test equipment.
12.102 In the case of non-rectangular canopies, an interpretation of the above strategy should be adopted that will yield a
[HTM 03-01 A – PDF page 149]
12.0 Acceptance testing: validation
When these tests have been completed, the source distribution head and sampling probe will be moved to the next test positions, working around the test grid in a clockwise direction.
12.107 The test stands will be positioned on each test point in turn and a pair of readings (challenge, then penetration) taken when the instrument has stabilised.
12.105 The sampling probe of the detector instrument is mounted on a test stand with its orifice facing outwards horizontally from the centre of the UCV canopy, 1 m above floor level. The sampling probe will be orientated at right angles to the partial wall when sampling along the sides of the test grid but will be set to bisect the angle when measuring at the corner test positions. (See Figure 4 for test locations and see photograph of entrainment test equipment on page 131.)
12.108 When taking a reading, the test person should stay within the UCV canopy footprint but on the side opposite the sampling probe.
12.109 A single measurement will be taken at the geometrical centre of the UCV canopy footprint. The centre measurement will be taken with the sampling probe mounted vertically 1 m above floor level.
12.106 The test will commence at the first test position, this being designated the left-most corner of the test grid when facing the wall housing the theatre control panel. The penetration will also be measured at the corresponding test point on the inner zone commencing at the corner nearest to the first test position.
Figure 4 Entrainment test locations for a 2.8 m ×2.8 m UCV terminal
Theatre control panel
Intake
3 4 + A + + + + + + + + +
B +
+ + + + + + + + +
Entrainment test isokinetic fishtail sampling head location and orientation, 1 m above floor level
x x x + x x x
+ +
x x x + x x x
C + D +
x x x + x x x
E +
+ + + + + +
x x x + x x x
F +
x x x + x x x
G +
Sampling head mounted vertically for centre reading
x x x + x x x
+ + + +
H +
+ I + + + + + + + + +
+ J + + + + + + + + +
Test grid layout is as for Figure 3. Entrainment test set-up and guidance is given in paragraphs 12.86 onwards.