HTM 03-01 A: _Hlk37000395 – Barn and semi-barn theatres / _Hlk37000395 – Hybrid theatres

Specialised ventilation for healthcare premises Part A: The concept, design, specification, installation and acceptance testing of healthcare ventilation systems · PDF pages 76–77 of 213 · Source file: HTM0301-PartA-accessible-F6.pdf
Unofficial text extract. Tables and figures may be incomplete. Authoritative version: official PDF from NHS England.

[HTM 03-01 A – PDF page 76]

should be resolved with the architect and VSG when considering the design:

theatre that it serves. A service corridor, if provided, is an ideal location. The control box should be clearly labelled with the identity of the operating theatre that it serves.

• In order to reduce the risk of pre- contamination, surgical instruments should not have to pass one operating position in order to get to the one that they are destined for.

Barn and semi-barn theatres

• Staff having scrubbed should not have to pass one operating position in order to get to the one that they are destined for.

• Waste material being cleared from an operating position should not pass another when being removed from the operating theatre.

8.115 There is no aerobiological reason why two or more UCV systems should not be installed in a common area if adequate spacing is provided. These are known as barn theatres and require special design considerations and operational discipline. The relative positions of the UCV units, temperature control range and location of doors and openings to other areas will all significantly affect the airflow at the operating positions.

• The operating positions and their UCV canopies should be placed in line and not staggered or offset, otherwise their airflow patterns will interfere with each other.

8.116 A barn theatre has two or more operating positions each ventilated by a UCV canopy all in one open operating theatre. There may be a common scrub trough, SPS preparation room and shared utility, all of which reduce the facility’s footprint. For reasons of privacy and dignity, there is usually a separate anaesthetic room for each operating position. The operating positions may be separated by glass screens to prevent bone fragments being propelled from one position to another when high-pressure air-driven surgical tools are being used.

8.117 A semi-barn theatre is very similar but would have a full-height dividing wall between the operating positions. The wall will not extend across the full width of the room, only its middle section. This creates a degree of physical separation between the operating positions but allows staff to walk from one to another around the ends of the dividing wall.

• While barns and semi-barns have, from the staffing and space utilisation point of view, many advantages, they can create problems with temperature control and energy efficiency. It is not advisable to shut their ventilation off at night or weekends as if the operating theatre temperature drops it can take a considerable time for the ventilation to achieve the required air velocity at the operating position (paragraph 8.96). Because the barn is a large open space, when it becomes cold the warm air supplied by the UCV canopies tends to rise and stratification occurs. As a result, although from the user’s point of view the ventilation appears to be running, the air being delivered does not actually have enough velocity to reach the operating table.

8.118 It is important that the physical layout and ventilation strategy of the barn or semi-barn are in harmony if the installation is to work successfully. The following points

• Access for service, maintenance and future upgrades or refurbishment will be restricted as this can only be carried out when none of the operating positions is in use.

[HTM 03-01 A – PDF page 77]

8.0 Specific healthcare department requirements

Neutral pressure theatres for infectious patients

8.119 Ventilation of each UCV canopy and associated anaesthetic room will be by a dedicated AHU; ventilation of the shared spaces and perhaps recovery area would either be shared between the operating position AHUs or provided by a separate AHU.

Hybrid theatres

8.124 The client may have a requirement for an operating suite for surgery on infectious patients. This may be a dedicated neutral-pressure operating suite or a standard operating suite that is designed to be easily convertible to a neutral-pressure suite. If airborne microorganisms liberated from a patient during a surgical procedure are allowed to cascade out into the adjacent corridors, they could infect other patients or the staff in the operating department.

8.125 The room provision and layout will be as for a conventional operating suite with the following variation to the ventilation scheme:

• The operating theatre will have a balanced supply and extract so that it is at the same pressure as the corridor.

8.120 A designation given to operating theatres that contain scanning equipment on a robotic arm. Major surgical procedures are carried out and the patient is scanned as necessary during the procedure. The scanning equipment may be floor-mounted or ceiling-hung and there will be one or more large monitors, a variety of screens and the medical gas terminal units all ceiling-hung on articulated pendants. The number of pendants and their supporting steelwork can reduce the space available to install ventilation ductwork and compromise the location of the supply air terminals. Liaison with the architects at an early stage in the project design is essential to ensure a satisfactory ventilation solution.

• Air should not cascade from the theatre to the surrounding rooms, so pressure stabilisers and/or transfer grilles should not be fitted. In the case of a convertible operating suite, permanently fitted hinge-down blanking plates with clamps should be provided to close the pressure stabiliser/transfer grille openings when required.

8.121 Hybrid theatres tend to be significantly larger than conventional theatres and may have a radiation protected control room and an equipment room to house the servers for the scanning equipment and its robotic arm in addition to the standard operating suite of rooms. The ventilation load will therefore be larger and standard solutions should be adapted to suit or the designer will need to return to first principles (see Appendix 8).

8.122 Because of the increased airflow requirement, the AHU will be larger than for a standard conventional operating suite.

• The preparation room may be dispensed with to avoid having stock that could become pre-contaminated. Sterile packs, instruments and consumables would be delivered to the theatre on a case by case basis. If a preparation room is required, it should be maintained at 10 Pa to both the theatre and corridor.

8.123 In all other respects the ventilation design and theatre control panel will be as for a conventional operating suite as above.

• The anaesthetic room should have a supply in excess of extract so that is maintained at 10 Pa above both the corridor and the theatre. There should be a pressure stabiliser between the